Introduction
The incidence of end-stage kidney disease (ESKD) in South Korea has more than doubled over the past 12 years, rising from 9,335 cases in 2010 to 18,598 cases in 2022. Additionally, the overall prevalence has increased by 90% over the past decade, from 70,211 cases in 2012 to 134,826 cases in 2022 [
1,
2]. This significant rise ranks among the highest globally, placing South Korea second in the incidence rate of ESKD among the member countries of the Organization for Economic Cooperation and Development [
3]. Hemodialysis remains the most widely utilized modality for kidney replacement therapy. While advancements in dialysis care and management have contributed to improved clinical outcomes [
1,
4,
5], the rising prevalence of ESKD and the growing reliance on hemodialysis have resulted in substantial economic burdens [
6]. Consequently, there is an increasing need to alleviate these burdens by implementing cost-effective strategies, including the domestic production of essential medical devices such as hemodialysis dialyzers.
Recently, the Synoflux series high-flux dialyzer (Synopex Inc.) received regulatory approval from the Korean Ministry of Food and Drug Safety as a class III medical device following a non-inferiority evaluation against commercially available dialyzers. The membrane is manufactured using a modified polyethersulfone (PES) membrane, which is known for its high biocompatibility, efficient toxin removal, and robust structural stability [
7,
8]. However, there is currently no clinical evidence available to verify its safety and efficacy in real-world patient settings. Therefore, this clinical trial aims to evaluate the non-inferiority of the domestically developed Synoflux series hemodialysis dialyzers compared to commercially available alternatives in Korean patients with ESKD. Additionally, this study seeks to collect and analyze safety data to ensure clinical applicability and regulatory compliance in routine hemodialysis practice.
Methods
Ethical considerations
This study received approval from the Institutional Review Boards of the participating clinical centers (see Additional information). All procedures conformed to the principles outlined in the Declaration of Helsinki and were conducted in accordance with the International Conference on Harmonization Good Clinical Practice guidelines. Written informed consent was obtained from all participants prior to enrollment, with additional safeguards implemented for elderly individuals as appropriate. Participant confidentiality was strictly maintained through the use of coded data and secure storage. Provisions for compensation and prompt management of adverse events were also established.
Study design
This study is a multicenter, randomized, crossover clinical trial designed to evaluate the non-inferiority of the Synoflux series dialyzer compared to the commercially available FX Classix dialyzer (Fresenius Medical Care AG & Co. KGaA) in patients with ESKD undergoing maintenance hemodialysis. After obtaining written informed consent (Visit 1), eligible participants underwent a screening period of up to 2 weeks to confirm their eligibility based on the inclusion and exclusion criteria. Qualified participants were then randomized in a 1:1 ratio into one of two groups: Group 1 (Synoflux dialyzer first) or Group 2 (control dialyzer first) at Visit 2. Randomization was performed using a block randomization method, with the allocation sequence generated by a central computer. This process was conducted by an independent statistician who was not involved in the clinical trial to ensure allocation concealment.
Patients underwent hemodialysis three times per week, with each session lasting 4 hours. The study comprised two consecutive 4-week treatment periods. In Period 1 (Visit 2 to Visit 3), Group 1 received the Synoflux dialyzer, while Group 2 received the control dialyzer. In Period 2 (Visit 4 to Visit 5), participants switched to the alternative dialyzer type, ensuring that all subjects were exposed to both devices. All dialyzers used in the study had a membrane surface area of either 1.4 or 1.8 m
2, and this surface area was consistently maintained after crossover to ensure comparability between treatment periods. The dialysis schedule and duration remained consistent across both groups. To ensure consistency in efficacy assessment, key dialysis parameters—including blood flow rate, dialysate flow rate, and session duration—were strictly standardized at 4 hours on the days of efficacy evaluation (Visit 3 and Visit 5). While ultrafiltration volume was individualized based on clinical needs, we ensured during study planning and execution that the variation in ultrafiltration between these visits would remain within 0.5 kg per participant to minimize confounding. The overall study design is summarized in
Fig. 1.
Eligibility criteria
Participants were eligible for inclusion if they met the following criteria: age ≥19 years; undergoing stable thrice-weekly maintenance hemodialysis for a minimum of 3 months prior to enrollment; provided written informed consent after receiving a comprehensive explanation of the study; and were deemed suitable for study participation through to completion at the discretion of the investigator.
Exclusion criteria included patients undergoing hemodialysis on a schedule other than three times per week; those receiving concomitant peritoneal dialysis; individuals planning to transition to peritoneal dialysis or undergo kidney transplantation during the study period; and those with residual renal function, defined as a daily urine output of ≥500 mL despite three-times-weekly dialysis. Patients were also excluded if they had a history of acute infection within 4 weeks prior to screening or were diagnosed with active malignancy, liver cirrhosis, HIV infection, bleeding disorders, severe congestive heart failure, recent myocardial infarction, stroke, seizure, thrombosis, or uncontrolled arrhythmia. Additionally, women who were pregnant, breastfeeding, or unwilling to use appropriate contraception were deemed ineligible for participation.
Efficacy endpoints
The primary efficacy endpoint was the urea reduction ratio (URR, %) at the 4th week of each dialysis period, calculated using the following formula: URR (%) = [(Cpre – Cpost) / Cpre] × 100, where Cpre and Cpost represent the pre- and post-dialysis blood urea nitrogen (BUN) concentrations, respectively. Single-pool Kt/V (spKt/V) was also assessed as an index of dialysis adequacy, calculated as follows: spKt/V = –ln (R – 0.008 × t) + [(4 – 3.5R) × UF/W], where R is the post-/pre-dialysis BUN ratio, t is the dialysis duration (hours), UF is the ultrafiltration volume (L), and W is the post-dialysis body weight (kg). Secondary efficacy endpoints included the reduction ratio (%) of middle-molecule solutes such as cystatin C, β2-microglobulin, myoglobin, prolactin, and interleukin-6 (IL-6).
Safety endpoints
The primary safety endpoint was the preservation of albumin, measured by tracking serum albumin levels, at three predefined time points: baseline (Visit 2), the crossover point (Visit 3), and the end of the study (Visit 5). Secondary safety assessments included the monitoring of adverse events during each dialysis session. Intradialytic hypotension (IDH) was defined as a nadir systolic blood pressure <90 mmHg following dialysis initiation. Dialyzer clotting was defined as the premature termination of dialysis due to circuit coagulation or the need for increased heparin dosing, and it was evaluated on a per-session basis. In addition to IDH, patient-reported symptoms, including dizziness, nausea, muscle cramps, and chest discomfort, were systematically collected during each dialysis session.
Statistical analysis
Between-group comparisons of continuous variables were conducted using either an independent t test or a Wilcoxon rank-sum test. Categorical variables were compared using the chi-square test or Fisher exact test, as appropriate. For paired comparisons of continuous variables between test and control dialyzers (e.g., spKt/V, URR, and reduction ratios of middle-molecule solutes), a paired t test or Wilcoxon signed-rank test was utilized based on the data distribution. Differences in outcomes were reported as mean differences with corresponding 95% CIs.
The primary efficacy endpoints were analyzed using a non-inferiority framework. Non-inferiority was assessed utilizing the two one-sided test procedures. The non-inferiority margin was predefined as 10% of the mean value of the control dialyzer for spKt/V and 5% for URR. For serum albumin, which served as the primary safety endpoint, a non-inferiority margin of 5% was applied to evaluate albumin preservation. Adverse events were summarized descriptively. IDH and dialyzer clotting events were assessed on a per-session basis, and formal hypothesis testing for adverse events was not the primary focus.
A priori sample size was calculated assuming a non-inferiority margin (δ) of 5%, within-subject standard deviation of 8.07%, and no true difference (ε = 0), based on parameters reported in a previous randomized controlled trial comparing a medium cut-off (MCO) dialyzer and a high-flux dialyzer [
9]. This yielded 21 patients per sequence for 80% power at a one-sided α of 2.5%. Accounting for a 15% dropout rate, the target enrollment was set at 50 patients.
All statistical analyses were conducted using SAS software, version 9.4 (SAS Institute Inc.). A two-sided p-value of <0.05 was considered statistically significant unless otherwise specified.
Discussion
This multicenter, randomized, crossover clinical trial evaluated the clinical performance and safety of a domestically developed dialyzer compared to a commercially available dialyzer in patients with ESKD. The results demonstrated that Synoflux met the predefined non-inferiority criteria for dialysis adequacy, as measured by URR and spKt/V when compared to the control dialyzer. Additionally, Synoflux exhibited comparable or superior clearance of middle-molecule solutes, particularly β2-microglobulin, myoglobin, and prolactin, while maintaining acceptable albumin retention within the predefined non-inferiority margin. The safety profile of both dialyzers was similar, with equivalent frequencies of adverse events and no serious dialyzer-related complications observed.
These findings align with previous studies investigating modified PES membranes, which have reported enhanced removal of middle molecules compared to conventional membranes. Specifically, one study demonstrated that as the molecular weight of the solute increased, the clearance advantage of the modified membrane became more pronounced, achieving a 124.8% higher reduction in myoglobin than a reference membrane [
10]. These performance improvements are attributed to the incorporation of negatively charged polyelectrolytes into the membrane surface during manufacturing, enhancing solute interaction without altering the base polymer composition. This structural modification results in a steeper sieving profile, facilitating efficient clearance of middle molecules while minimizing albumin loss [
7]. In another study, the modified PES membrane exhibited comparable removal of small solutes and low-molecular-weight proteins relative to online post-dilution hemodiafiltration using conventional high-flux membranes while providing superior clearance of myoglobin compared to a polyacrylonitrile-based membrane [
11].
In addition to optimizing sieving performance, membrane development has also concentrated on enhancing fiber geometry to improve solute transport. Reducing the diameter and wall thickness of hollow fibers decreases the diffusion path and facilitates diffusive clearance, while larger molecules necessitate convective transport for effective removal [
12]. The membrane evaluated in our study demonstrated advantageous structural characteristics, featuring a mean pore size of approximately 8 nm—double that of its comparator—and a wall thickness of 30 µm, which is about 10% thinner than that of commonly used comparator membranes. These features enhance diffusive and convective solute transport, particularly improving the clearance of middle molecules. Additionally, the dialyzer incorporates spacer yarns—fine threads strategically positioned between hollow fibers—to maintain consistent inter-fiber spacing and optimize dialysate flow distribution. Previous studies have demonstrated that spacer yarns reduce flow channeling and dead space within the fiber bundle, thereby improving solute clearance. They also promote uniform dialysate flow and minimize surface fouling, which together improve dialyzer performance and biocompatibility [
13,
14]. The inclusion of this design element in the test dialyzer may have contributed to the favorable clearance outcomes observed in our study.
The enhanced removal of middle molecules is particularly significant due to the established association of these toxins with dialysis-related complications, including amyloidosis, cardiovascular disease, and chronic inflammation. β2-microglobulin, which has a molecular weight of 11.8 kDa, is a well-recognized contributor to dialysis-related amyloidosis and is linked to long-term musculoskeletal complications such as carpal tunnel syndrome and destructive arthropathy [
15,
16]. Enhanced clearance of β2-microglobulin has been associated with improved functional status, as well as better cardiovascular and overall survival outcomes in dialysis patients [
17,
18]. Cystatin C (13.3 kDa), traditionally utilized as a biomarker for glomerular filtration rate, also reflects systemic inflammation and has been independently associated with increased arterial stiffness, left ventricular hypertrophy, and adverse cardiovascular events in dialysis patients [
19–
22]. The effective removal of cystatin C may help reduce the cardiovascular risk burden in this population. Prolactin (23 kDa), often elevated in ESKD due to reduced renal clearance, is linked to multiple pathophysiological consequences, including gonadal dysfunction, menstrual irregularities, sexual dysfunction, and immune dysregulation [
23–
26]. Its successful clearance may provide therapeutic benefits in addressing these frequently overlooked complications of ESKD.
In this study, IL-6 levels did not significantly decrease following dialysis with either dialyzer, reflecting the known limitations of conventional high-flux membranes in removing cytokines of relatively large molecular weight (~26 kDa) and partial protein-binding characteristics. This observation is consistent with prior evidence suggesting that IL-6 is poorly cleared by diffusion- or convection-based modalities. Notably, IL-6 levels also did not show a significant post-dialysis increase, suggesting that neither membrane provoked a measurable pro-inflammatory response during treatment. While these findings do not allow definitive conclusions regarding biocompatibility, they are reassuring from an immunologic standpoint.
Both dialyzers demonstrated favorable safety profiles, characterized by preserved serum albumin levels and low incidences of IDH, clotting events, and other serious dialysis-related complications. A previous study evaluating the safety of MCO membranes confirmed modest albumin loss; however, these reductions have not been associated with clinically significant consequences or the need for albumin supplementation [
27]. In the present study, although a mild decrease in serum albumin was observed in one group, the reduction remained within the predefined non-inferiority margin and did not result in a statistically significant difference across the overall study population.
This study has several limitations. Firstly, the short-term crossover design inherently restricts the assessment of long-term outcomes, such as inflammation, hospitalization, and mortality. Secondly, this study did not include a formal wash-out period between treatment phases. However, in maintenance hemodialysis patients, the physiological and biochemical milieu is effectively reset during the inter-dialytic interval, which may mitigate the need for an extended wash-out period. Thirdly, although a crossover design was employed, statistical correction for potential carry-over effects was not conducted.
However, considering the relatively short duration of each treatment period and the clinical stability of the enrolled patients, the risk of significant carry-over effects is likely minimal. Additionally, Synoflux was compared with a single, widely used imported high-flux dialyzer rather than multiple control membranes. A more comprehensive comparison may be necessary to fully elucidate the clinical versatility of the Synoflux dialyzer. Future studies should incorporate biomarkers of inflammation, oxidative stress, and patient-reported outcomes to better characterize the biological impact and patient-centered effects of this novel PES membrane technology.
In conclusion, the Synoflux dialyzer demonstrated clinical non-inferiority in terms of dialysis adequacy and safety when compared to an imported standard, with potential advantages in middle-molecule clearance. These findings, combined with the structural and mechanical benefits of yarn-based hollow fiber membranes, support the feasibility of domestic production and broader utilization of next-generation dialyzers in Korea. Ongoing clinical evaluations and further industrial optimization are essential to fully realize the clinical and economic potential of this technology.